T1 - Investigation of alanine dosimeter for output measurement in radiotherapy
T2 - Results from multicenter in Thailand
AU - Intang, Aungsumalin
AU - Oonsiri, Puntiwa
AU - Kingkaew, Sakda
AU - Chatchumnan, Nichakan
AU - Tangboonduangjit, Puangpen
AU - Stansook, Nauljun
AU - Changkaew, Pimolpun
AU - Khachonkham, Suphalak
AU - Funsian, Amporn
AU - Rattanaareeyakorn, Warinthorn
AU - Turathong, Suriyaporn
AU - Krongkietlearts, Komkrit
AU - Seenukhroah, Adcharee
AU - Chaisuttee, Rusdchai
AU - Sirichotwatcharakit, Chompunek
AU - Oonsiri, Sornjarod
N1 - Publisher Copyright:
© 2023, Faculty of Associated Medical Sciences, Chiang Mai University. All rights reserved.
PY - 2023/5/1
Y1 - 2023/5/1
N2 - Background: Alanine dosimeters are generally used in high-dose industrial applications (kGy). Later, research into employing alanine as a dosimeter in radiotherapy (1-20 Gy) has increased, since alanine may be an alternative transfer dosimeter for quality control, postal audit, and intercomparison between laboratories. However, several factors such as the dosimeter’s characteristic should be investigated while utilizing alanine in radiotherapy. In addition, the optimal electron paramagnetic resonance (EPR) reader should be configured to match the absorbed dose range. Objectives: This study aims to optimize the EPR setting parameters, study the characteristics of alanine dosimeters, and estimate the uncertainty of the 6MV-FFF linear accelerator in a dose ranging from 1 to 20 Gy. The output measurements from different facilities were also investigated. Materials and methods: The alanine dosimeters were irradiated with a 6MV-FFF beam using linear accelerator, Varian TrueBeam (Varian Medical Systems, Inc, CA, USA), 100 cm SSD, with a field size of 10x10 cm2 at 1 to 30 Gy. The EPR operation parameter has been optimized for these dose ranges. The characteristics of alanine dosimeters were then investigated, along with the estimation of uncertainty in using alanine. Finally, the alanine dosimeter proficiency was validated using 9 distinct linear accelerator machines. Results: The EPR parameters were found to be optimized at 1.589 mW of MP, 7.018 G of MA, and 40.96 ms of TC. The expanded uncertainty (k=2) was reported at 2.68% in the 1-20 Gy dose range. The alanine dosimeters’ characteristics were found to have good uniformity and reproducibility, low fading, and angle-and dose-independence. Although the investigation was performed in 9 different linear accelerator machines, the difference of delivered dose output was measured and reported with difference percentages within ±1%. Conclusion: This study reports the feasibility of using alanine dosimeters in radiotherapy. The important EPR parameter setting, and alanine dosimetry characteristics were investigated, whose results suggest that alanine can be used at a dose range of 1-30 Gy. This especially benefits the SRS treatment which uses a high dose per fraction, and this dosimeter can be an alternative transfer dosimeter. Nonetheless, various factors should be explored using an appropriate phantom prior to clinical application.
AB - Background: Alanine dosimeters are generally used in high-dose industrial applications (kGy). Later, research into employing alanine as a dosimeter in radiotherapy (1-20 Gy) has increased, since alanine may be an alternative transfer dosimeter for quality control, postal audit, and intercomparison between laboratories. However, several factors such as the dosimeter’s characteristic should be investigated while utilizing alanine in radiotherapy. In addition, the optimal electron paramagnetic resonance (EPR) reader should be configured to match the absorbed dose range. Objectives: This study aims to optimize the EPR setting parameters, study the characteristics of alanine dosimeters, and estimate the uncertainty of the 6MV-FFF linear accelerator in a dose ranging from 1 to 20 Gy. The output measurements from different facilities were also investigated. Materials and methods: The alanine dosimeters were irradiated with a 6MV-FFF beam using linear accelerator, Varian TrueBeam (Varian Medical Systems, Inc, CA, USA), 100 cm SSD, with a field size of 10x10 cm2 at 1 to 30 Gy. The EPR operation parameter has been optimized for these dose ranges. The characteristics of alanine dosimeters were then investigated, along with the estimation of uncertainty in using alanine. Finally, the alanine dosimeter proficiency was validated using 9 distinct linear accelerator machines. Results: The EPR parameters were found to be optimized at 1.589 mW of MP, 7.018 G of MA, and 40.96 ms of TC. The expanded uncertainty (k=2) was reported at 2.68% in the 1-20 Gy dose range. The alanine dosimeters’ characteristics were found to have good uniformity and reproducibility, low fading, and angle-and dose-independence. Although the investigation was performed in 9 different linear accelerator machines, the difference of delivered dose output was measured and reported with difference percentages within ±1%. Conclusion: This study reports the feasibility of using alanine dosimeters in radiotherapy. The important EPR parameter setting, and alanine dosimetry characteristics were investigated, whose results suggest that alanine can be used at a dose range of 1-30 Gy. This especially benefits the SRS treatment which uses a high dose per fraction, and this dosimeter can be an alternative transfer dosimeter. Nonetheless, various factors should be explored using an appropriate phantom prior to clinical application.
KW - Alanine dosimeter
KW - output measurement
KW - radiotherapy
UR - http://www.scopus.com/inward/record.url?scp=85150160539&partnerID=8YFLogxK
U2 - 10.12982/JAMS.2023.039
DO - 10.12982/JAMS.2023.039
M3 - Article
AN - SCOPUS:85150160539
SN - 2539-6056
VL - 56
SP - 106
EP - 116
JO - Journal of Associated Medical Sciences